Performance verification of ABI Quantistudio Dx real-time fluorescent quantitative PCR HBV-DNA detection system

LI Pengfei, GAO Feng, and ZHANG Yanhui

Medical Journal of the Chinese People Armed Police Forces ›› 2019, Vol. 30 ›› Issue (2) : 109-112.

PDF(738 KB)
PDF(738 KB)
Medical Journal of the Chinese People Armed Police Forces ›› 2019, Vol. 30 ›› Issue (2) : 109-112.
ORIGINAL ARTICLES

Performance verification of ABI Quantistudio Dx real-time fluorescent quantitative PCR HBV-DNA detection system

  • LI Pengfei1, GAO Feng1, and ZHANG Yanhui2
Author information +
History +

Abstract

Objective To verify the performance of ABI Quantistudio Dx Real-time PCR HBV-DNA detection system.Methods According to CNAS-CL36 and GB/T 22576-2008, the analytical performance, such as precision, accuracy, limit of detection and linearity,was determined by ABI Quantistudio Dx real-time PCR system. Patient specimens, quality products, and reference materials from the National Center for Clinical Laboratories (NCCL) were employed for performance verification.Results The coefficient of variation (CV) of intra-assay of specimens with low concentration and high concentration of ABI Quantistudio Dx real-time PCR system was 2.09% and 1.3%, respectively, while the CV of inter-assay of specimens with a low concentration and high concentration was 3.92% and 2.88%, respectively. Analysis of reference materials from NCCL showed that test results were highly accurate. The detection limit of this system was 50 IU/ml. The linear regression equation was Y=1.0001X, R2=0.9976. The linearity ranged from 1.00×102 to 1.00×108 U/ml.Conclusions The performance of ABI Quantistudio Dx Real-time PCR HBV-DNA detection system meets the needs of clinical detection.

Key words

performance verification / real-time quantification PCR / hepatitis B virus DNA quantification

Cite this article

Download Citations
LI Pengfei, GAO Feng, and ZHANG Yanhui. Performance verification of ABI Quantistudio Dx real-time fluorescent quantitative PCR HBV-DNA detection system[J]. Medical Journal of the Chinese People Armed Police Forces. 2019, 30(2): 109-112

References

[1] 邢卉春, 庞 婷. 从2015年中国《慢性乙型肝炎防治指南》看乙肝防治研究动态[J].武警医学, 2016,27(5):433-436.
[2] 中国国家标准化管理委员会. GB/T22576-2008医学实验室质量和能力的专用要求[S]. 北京: 中国国家标准化管理委员会, 2010.
[3] 中国合格评定国家认可委员会. CNAS-CL02-A009:医学实验室质量和能力认可准则在分子诊断领域的应用说明[S]. 北京:中国合格评定国家认可委员会, 2018.
[4] Clinical and laboratory standards institute. EP15-A2: User verification of performance for precision and trueness: approved guideline-second edition[S].Wayne, PA, USA: NCCLS, 2005.
[5] Clinical and laboratory standards institute. EP6-A: Evaluation of the linearity of quantitative measurement procedures: a statistical approach, approved guideline[S]. Wayne, PA, USA: NCCLS, 2003.
[6] 辛 娜, 杨 超, 白跳艳, 等. 自建 HBV-DNA 实时荧光定量 PCR检测系统性能验证方法及结果分析[J]. 山西医科大学学报, 2016,47(4):365-368.
[7] Li M, Chen L, Liu L M, et al. Performance verification and comparison of Tian Long automatic hypersensitive hepatitis B virus DNA quantification system with roche CAP/CTM system [J]. World J Gastroenterol, 2017, 23(37):6845-6853.
[8] 王 娇, 钱 军, 石 英, 等. Mx3000pTM实时荧光定量PCR仪HBV-DNA检测系统的性能验证 [J].中国医药导报, 2018, 15(15): 65-68.
[9] 俞 刚, 陈 瑜, 张 佳, 等. 罗氏Light Cycler 480实时荧光定量PCR系统检测HBV-DNA性能验证 [J]. 标记免疫分析与临床, 2016, 23(10):1215-1225.
[10] 巢 薇. ABI 7300实时荧光定量PCR仪的性能评估 [J]. 国际检验医学杂志, 2015, 36(4): 494-495.
[11] 胥明勇, 朱华强. ABI 7500 HBV-DNA检测系统性能评价 [J]. 实验与检验医学, 2017, 35(4):489-491.
[12] 赵 芹, 辛青松, 李峥嵘, 等. ABI ViiA 7 Taqman HBV-DNA检测系统性能验证 [J]. 国际检验医学杂志, 2015, 36(5):669-671.
[13] 杨佳佳, 张彦懿, 刘华伟, 等. HBV-DNA实时荧光定量PCR检测试剂盒性能验证[J]. 国际检验医学杂志, 2018, 39(2):213-217.
[14] 高 峰, 高慧华, 杨学文. 脂血因素对荧光定量PCR检测乙肝病毒DNA的影响及解决措施[J]. 临床检验杂志, 2007, 25(5):359-360.
[15] Killeen A A, Long T, Souers R, et al. Verifying performance characteristics of quantitative analyticalsystems: calibration verification, linearity, and analytical measurement range[J]. Arch Pathol Lab Med, 2014, 138(9):1173-1181.